SECTION THREE
3-2
Tab Down Tab Up Enter Back Read On/Off
Active Illuminate/Observer
1. Press Tab Up or Tab Down to move the highlight to Active Illum/Obs and press
Enter to open the Active Illum/Obs editor.
2. Press Tab Up or Tab Down to move the highlight to the desired 10° combination and
press Enter to toggle the combination active or inactive. The arrow (>) indicates
active.
3. To edit 2° combinations, highlight To 2° using Tab Up or Tab Down and press Enter.
4. Press Tab Up or Tab Down to move the highlight to the desired 2° combination and
press Enter to toggle the combination active or inactive. The arrow (>) indicates
active.
5. After edits are complete, press Back to save and exit the illum/obs editor.
Active Angles
1. Press Tab Up or Tab Down to move the highlight to Active Angles and press Enter
to open the Active Angles editor.
2. Press Tab Up or Tab Down to move the highlight to Angle #1 and press Enter to
toggle between 15° and 25°.
3. Press Tab Up or Tab Down to move the highlight to Angle #3 and press Enter to
toggle between 75° and 110°.
4. After edits are complete, press Back to save and exit the editor.
Measure Options
The Measure option allows you to configure the following settings:
Auto Ref – Sets the automatic reference option. When Auto Reference is enabled (on),
the stored reference that most closely matches the measured sample is automatically
selected during a Store or Normal measurement. When Auto Reference is disabled (off),
you must manually select a reference before taking a measurement.
Averaging –Sets the number of readings (0-99) required to obtain an averaged
measurement.
Read Action – Allows you to have the measurement activated by bottom switches (Foot
Switches only), Read button (Read Key only), or in combination of the Read button and
bottom switches (Switch & Key req.).
Editing Measure Options
1. Press Tab Up or Tab Down to move the highlight to Measure Options.
CONFIGURATION
Display Options
Measure Options
Set Clock
Communications
Language
2. Press Enter to access the Measure Options editor.
Measure Options
Auto Ref: On
Averaging : 1
Read Action