SERVICE AND GENERAL MAINTENANCE
5-3
Troubleshooting Tips
Problem Cause Solution
The instrument display is
hard to read.
Screen contrast set
incorrectly.
Adjust the contrast as explained
in Section One.
AC adapter is not
attached.
Check the connection on the
back of the docking station and
at the AC adapter.
The instrument’s battery
pack will not charge.
Internal battery pack is
bad.
Contact an authorized service
center.
Instrument is not properly
seated in the docking
station.
Make sure instrument is
properly seated over the
contacts.
The interface cable is not
attached.
Check the connection on the
back of the docking station and
at the computer.
Measurements not uploading
to computer when the
instrument is seating in the
docking station.
Contaminated connectors. Disconnect docking station
from computer and wall outlet.
Clean connectors on the
instrument and the docking
station with a cotton swab
dampened with Isopropyl
alcohol. Cotton swab should
not be saturated.
Calibration reference is
dirty.
Clean the calibration reference
as specified in section 5.
Instrument will not calibrate
correctly.
Optics is dirty. Clean the optics as specified in
section 5.
Instrument is in power
down mode.
Press the Power button to
active the instrument.
Screen contrast is set too
low.
Adjust the contrast as explained
in Section One.
The instrument screen does
not display anything.
Internal batteries require
charging.
Place instrument on docking
station and charge battery pack
for a minimum of 4 hours
before use.
The optional barcode reader
does not work.
Instrument is seating in
the docking station.
Remove the instrument from
the docking station. Barcode
reader is disabled in the
docking station.