Calibration Mode
You only need to calibrate the instrument if the measurement performance is in question, if it has
been dropped or if it received an excessive jarring. A calibration is completed by measuring a
variety of patches on the calibration card.
NOTE: Make sure you use the calibration card that was supplied with the instrument for
calibrating. Do not substitute a calibration card from another instrument. The serial number on
the card should match the cal plaque number displayed in the Info screen.
Calibrating the Instrument
1. From the Main screen, use the Navigation Ring to move the highlight focus to the Calibration
icon.
or (“X” on the graphic indicates a verification failure)
2. Press the Select button.
The last calibration date and time appears. To exit calibration mode without calibrating, select
the Exit icon.
3. Activate targeting mode by pressing the Measure button; or press the Select button while
the Begin Arrow icon is highlighted.
4. Remove the calibration card from its envelope and position the instrument’s target window on
the white patch. Use the reticule icon in the display screen for positioning. Make sure the
entire instrument is positioned on the calibration card as shown below.
Begin arrow
Date and time of last calibration
Cal reference serial number
White patch
Calibration
mode icon
IMPORTANT: This measurement
calibrates and stores images for
each illumination and may take up
to 60 seconds per illumination to
complete. Do not move or bump
the instrument during this time or
the measurement could fail. The
screen will display which
illuminate is being calculated.