INSTRUMENT FUNCTIONS
5-47
To measure a reference:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference. The reference values
displayed are the current values set in the instrument.
NOTE: Reference will not appear in the measurement list unless
PAPER INDIES–REF or REFLECT INDIES—REF is selected as the
active paper indice mode. Refer to Paper Indices Measurement
Mode earlier in this section.
2. Center the instrument target window over the paper
reference and lower the instrument to take a reading.
The reference values are updated.
To manually edit reference values:
1. Select PAPER INDICES—REF or REFLECT
INDICES—REF measurement mode.
2. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference. The reference values
displayed are the current values set in the instrument.
NOTE: Reference will not appear in the measurement list unless
PAPER INDIES–REF or REFLECT INDIES—REF is selected as the
active paper indice mode. Refer to Paper Indices Measurement
Mode earlier in this section.
3. Press the Enter key # to move the highlight to the
data side of the screen. <
Enter Ref> appears in the
user dialog.
NOTE: Highlighting Clear Ref and pressing Enter key # can
quickly clear current reference data.
4. Press the Tab Up key $ or Tab Down key @ to
highlight desired attribute.
5. Press the Enter key # to access
References menu.
6. Use the Tab keys $@ to move the cursor to desired
digit. Press Enter key # to access editor.
PAPER INDICES—REF
Sample
Mean
Reference
Br 85%
Ct 2%
Clear Ref
<Measure Ref>