CHAPTER FIVE
5-20
Measuring References
The reference function is used to measure reference data into a
selected group. A total of 1,424 references can be stored in the
instrument. The default reference name that is generated after
a measurement (i.e., Ref0001) can be change to a specific
color name or code.
NOTE: Reference databases that are downloaded using ColorMail
Express cannot be edited or deleted. The word <LOCKED> appears
is the user dialog, indicating the reference cannot be changed.
MATCH:Group1 Options
Sample
Match
L* 34.94
a* —1.98
b* —25.65
<View Refs>
D50/2
To create a reference:
1. Make sure that the active group you want to store the
reference in is displayed (see Setting Options for
selection procedure).
2. Press the Tab Up key $ or Tab Down key @ to
highlight References and press the Enter key #.
REF:Group1 Options
Add New
<Measure Ref>
D50/2
3. Make sure Add New is highlighted. Center the target
window over the reference and lower the instrument to
take a reading.
REF:Group1 Options
Add New
Ref0001
Ref0001
L* 33.44
a* –1.27
b* –29.92
Delete
<View Refs>
D50/2
The instrument automatically numbers (Ref0001)
each reference by default.
References