X-Rite 504 SpectroDensitometer Operator's Manual Page 29

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 28
USER INTERFACE
3-3
Full Calibration
The Full Calibration function is used to update the white
and black calibration points in the instrument.
To update calibration:
1. Repeatedly press the Tab Down key @ to highlight the
Configuration option.
2. Press the Enter # key. The
CONFIGURATION menu opens.
3. Use the Tab keys $@ to highlight Cal Options.
4. Press the Enter # key to open the
Calibration
Options
menu.
5. Use the Tab keys $@ to highlight Full
Calibration.
Calibration Options
Full Calibration
Enter Reflectances
Cal Alert: 24 hrs
<Select Options>
MAIN MENU
Hue/Graynes
Paper Indic
Compare
Calibration
Configuration
English
9600 bd
On bpr
X.XX V
CONFIGURATION
Language :English
Active Functions...
Density Opt :T
Color Option:CIE
Cal Options :24 hrs
Option Settings
Options
CONFIGURATION
Language :English
Active Functions...
Density Opt :T
Color Option:CIE
Cal Options :24 hrs