CHAPTER FIVE
5-28
Measuring Paper
The instrument takes the density value of the paper and
automatically subtracts it from subsequent solid and dot
measurements. The paper measurement values are
applied to all functions that support minus paper.
To measure paper:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Paper. <Measure Paper>
appears in the user dialog. The paper values
displayed are the current values set in the instrument.
2. Center the instrument target window over a sample of
the paper, and lower the instrument to take a reading.
3. The paper values are updated and the display
highlight advances to Solid.
NOTE: Press the Enter key # when Paper is highlighted to view the
RGBO values if HiFi status is used.
Measuring Dot Procedure
So far, you have performed the procedures to select the dot
mode, options, and to measure paper.
You are now ready to begin taking dot measurements to
evaluate dot gain and dot area. The measurement data that
displays depends on the way you set up your instrument
earlier, and the dot mode selected. The instrument
maintains separate data for each of the colors (vcmyrgbo).
DOT AREA Options
Paper
Solid
Tint
DOT AREA
V 0.09
C 0.08
M 0.09
Y 0.10
<Measure Paper>
T
DOT AREA Options
Paper
Solid
Tint
DOT AREA
M 0.00
<Measure Solid>
T
Page view 81