X-Rite 504 SpectroDensitometer Operator's Manual Page 89

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 88
INSTRUMENT FUNCTIONS
5-35
4. Lower unit to target window and hold closed. Once
measurement data is displayed, release the unit.
5. Make sure Ink1 is highlighted on the screen and
center the target window on first ink down patch.
6. Lower unit to target window and hold closed. Once
measurement data is displayed, release the unit.
Measurement data appears as actual data or difference
data. Refer to Viewing Trap Data later in this section.
Saving Trap Data as a Reference
The last trap measurement taken can be saved as a
reference. This allows you to compare subsequent trap
measurements and view the data difference. Trap data can
also be manually edited if desired.
To save a trap measurement as a reference:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference.
NOTE: Reference will not appear in the measurement list unless
TRAP–REF is selected as the active trap mode. Refer to Trap
Measurement Mode earlier in this section.
2. Press the Enter key # to access the References
Option screen.
3. Use the $@ keys to highlight the Store option and press
Enter key #. The trap percentage value is displayed and
color over color designation is now saved as the reference.
To manually edit trap reference values:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference.
NOTE: Reference will not appear in the measurement list unless
TRAP–REF is selected as the active trap mode. Refer to Trap
Measure Mode earlier in this section.
TRAP-REF
Options
Ink2
Ink1
Trap
Reference
C
Y
91%
<Modify Ref>
T
Page view 88
1 2 ... 84 85 86 87 88 89 90 91 92 93 94 ... 125 126